Low Perfusion and Missed Diagnosis of Hypoxemia by Pulse Oximetry in Darkly Pigmented Skin: A Prospective Study

Maureen F. McLaughlin, MS, RN, ACNS-BC, CPAN, CAPA, FASPAN

Originally published in Mayflower, the official newsletter of the Massachusetts Society of Perianesthesia Nurses (MASPAN).

Background: The use of pulse oximetry has been a standard of care in many clinical settings, including the post anesthesia care unit (PACU) and all anesthetizing locations. Developed in the 1970’s by Dr. Takuo Aoyagi, a Japanese engineer, a pulse oximeter estimates the oxygen saturation of blood in the peripheral circulation through the use of a light-emitting probe that measures oxygenated and deoxygenated blood. It was approved for medical use in the early 1980’s. Validation studies for medical devices such as the pulse oximetry are typically performed on healthy young adult volunteers with minimal comorbidities. Accuracy of oximeter performance may be impacted by skin melanin, anemia, motion, low perfusion and dyshemoglobins such as carboxyhemoglobin, when carbon monoxide binds with hemoglobin. Instances of occult hypoxemia were identified during the pandemic in which patients with darker skin pigmentation maintained adequate saturations measured by pulse oximetry but in fact had hypoxemia as evidenced by arterial blood gas analysis. Delayed identification and treatment of hypoxemia may result in patient harm. 

Abstract: The purpose of this research study was to identify whether the pulse oximeter overestimates oxygen saturation in the setting of increased skin melanin and low perfusion. The study used 146 healthy nonsmoking volunteers with varying skin tones determined using the Fitzpatrick scale. Informed consent and Institutional Review Board approval were obtained. 

Methods: Participants were placed in a semi recumbent position and an arterial line was inserted in order to obtain arterial blood gas analysis. As part of the hypoxemia protocol, participants inhaled air-nitrogen-carbon dioxide (CO2) mixtures via a mouthpiece and partial rebreathing system to achieve multiple levels of arterial oxygenation and relatively constant end-tidal CO2 levels. Stable saturations between 60% and 100% were targeted by the provider overseeing the inhaled mixture. At each level of saturation, 2 separate arterial blood gas samples were collected, each approximately 30 seconds apart. Pulse oximetry readings that experienced signal failure or failure to reach a planned level were discarded from the study. Researchers analyzed the correlation between skin pigment (light, medium or dark), perfusion index (saturation by pulse oximetry) and degree of hypoxemia. 

Using a multivariate mixed-effects analysis, the primary outcome findings of this study identified the correlation between low perfusion states, medium and dark skin pigmentation and degree of hypoxemia contributed to occult hypoxemia: false normal saturations measures by pulse oximetry contrasted to arterial blood gas sampling confirming  much lower values. 

Discussion: Missed diagnosis of hypoxemia occurred in 8% to 30% of readings in which the pulse oximeter indicates a saturation of 92% to 96% compared to arterial blood gas sampling results < 88% in patients with medium and dark skin pigmentation. Participants with medium skin pigmentation self-identified their ethnicity as Asian, Hispanic or Multiethnic. The researchers conclude that dark skin pigmentation combined with low perfusion is associated with missed identification of hypoxemia by pulse oximetry. 

Comments: Pulse oximetry has been a mainstay in anesthesia and perianesthesia nursing for decades. The pandemic highlighted concerns of occult hypoxemia associated with darker skin pigmentations. This research study, performed at the Hypoxia Research Laboratory at the University of Southern California, aligned low perfusion states and skin pigmentation with altered oxygen saturation using both continuous pulse oximetry and arterial blood gas analysis. Further research is underway by these researchers. They issue a strong recommendation that pulse oximeter regulatory guidance must reflect varying skin tones in the setting of low perfusion. 

Gudelunas, M.K., Lipnick, M., Hendrickson, C., Vanderburg, S., Okunlola, B., Auchus, I. et. al. (2024). Anesthesia and Analgesia, 138(3), 552-561.